CRAL150 | CRAL300 | CRAL300R | CRAL400 | CRAL400R | CRAL600 | CRAL800 | Custom | Components
Cool Laser Optics | Warranty

Cool Laser Optics Products by CustomChill

 
CustomChill, Inc. is licensed to manufacture and sell the patented Cool Laser Cutaneous Cooling Devices. This complete solution consists of the Cool Laser Optics (CLO®) devices with the dependable Custom Chill chillers.
  • Slim Pack
  • Cool Wand
  • Chiller

Cutaneous Cooling Devices
During laser treatments it may be desirable to limit injury to only certain aspects of the skin. Other than choosing appropriate wavelength and pulse duration, the option of adjunctive cooling while lasing has been used with a method for enhancing selective injury. This method is now used for vascular lesions, hair removal, and collagen restructuring. The Cool Laser Optic devices you will see elsewhere on this web-site may be used with any cosmetic lasers in the visible or near-infra-red range, and are ideally suited to meet clinical requirements for enhancing selective injury.

Proven Technology.
Simultaneous and continuous cooling during laser treatment reduces collateral thermal injury while maintaining selective injury of the targeted tissue. Cool Laser Optics technology was developed by a laser dermatologist, and is used by major laser manufacturers. It is FDA cleared for marketing.

Proven Effective.
Cool Laser Optics products allow you to use the higher laser energies often needed for effective treatment of a variety of conditions, while at the same time maintaining patient comfort. The results are irrefutable.

Proven Economical.
If you presently use cosmetic lasers, this cooling system is by far the most cost-effective way to expand their applications. If you are contemplating purchasing a cosmetic system, consider the economic advantages of acquiring stand-alone cooling.

Product Descriptions:

Cool Kit - The ideal configuration of cooling devices for the laser expert or novice. The Cool Kit contains the cooling devices you need for effective and efficient treatment. Included in the Cool Kit are the Recirculating Chiller, Slim Pack™ and Cool Wand™, along with a 12' hose assembly, carrying case and extra connectors.
CLO® Slim Pack™ - Lightweight, durable and easy to use, the Slim Pack is superb for cooling large areas prior to, during, and after treatment. This device provides excellent visualization of the target, while its large (4"x2") cooling footprint increases treatment efficiency. The Slim Pack is used with the CustomChill Recirculating Solid State Chiller. Applications of Cool Laser Slim Pack  by CustomChill, Inc.
CLO® Cool Wand™ - Flexible and maneuverable, the Cool Wand is ideal for small, hard to get at targets. This easy to use device is the perfect complement to the Slim Pack™. The Cool Wand™ is used with the CustomChill Recirculating Solid State Chiller. Applications of Cool Laser Wand by CustomChill, Inc.
CustomChill Recirculating Solid State Chiller - Reliability and precise temperature control in a small, lightweight and environmentally friendly package. When coupled with either a Slim Pack™ or Cool Wand™, this efficient, solid state chiller takes you where you need to go for effective skin cooling. Uses only distilled water to achieve the desired cooling effect. CRAL300 - 300 Watt Free-Standing Chiller with front outlet by CustomChill, Inc

Indications and Treatment Parameters for CLO® Cutaneous Cooling Device

Indications:
Use of CLO® Cutaneous Cooling Devices may be indicated in applications in which patients present intolerance to the pain associated with standard laser treatment.

Laser Compatibility:
CLO® devices can be used with lasers and intense pulsed light sources emitting visible or near-infrared wavelengths, including 488nm, 514nm, 532nm, 577nm, 585nm, 590nm, 595nm, 600nm, 694nm, 755nm, 800nm, 810nm, and 1064nm. The devices are not compatible with high peak power Q-Switched systems, Erbium, or CO2 lasers.


Treatment Parameters:

Fluence and Power Settings

The following are the recommended levels of energy and power settings for the specified lasers when using CLO® devices:

Laser Type Usual Fluence w/o CLO® Recommended Fluence w/CLO®* Usual Power Setting w/o CLO® Recommended Power Setting w/CLO®*
Argon
(488/514nm)
    1 - 2 Watts 3 - 4 Watts
Argon Pumped Turnable Dye
(577nm/585nm)
    0.7 - 0.8 Watts 1.5 Watts
KTP (CW)
(532nm)
    0.7 - 0.8 Watts 6 - 7 Watts
KTP (ms pulse)
(532nm)
8-12 J/cm2 10-15 J/cm2    
Copper Bromide
(578nm)
20-30 J/cm2 40-55 J/cm2    
Pulsed Dye
(585nm)
3-7 J/cm2 7-11 J/cm2    
Pulsed Dye
(590-600nm)
8-15 J/cm2 15-23 J/cm2    
LPIR(Alex.)
(755nm)
15-50 J/cm2 15-50 J/cm2    
Ruby (Non-Q-Switched)
(694nm)
8-12 J/cm2 10-20 J/cm2    
LPIR(Yag.)
(1064nm)
75-150 J/cm2 90-175 J/cm2    


Spot Sizes and Pulse Duration

The following are the recommended spot sizes and pulse durations for the specified lasers when using CLO® devices:

Laser Type Recommended Spot Size Recommended Pulse Direction
Argon
(488/514nm)
0.5mm* 0.5 Seconds
Argon Pumped Turnable Dye
(577nm/585nm)
0.6mm* 0.5 Seconds or Continuous Beam #
KTP (CW)
(532nm)
1.0mm** 0.5 Seconds
KTP (ms pulse)
(532nm)
1.0 - 5.0mm 10-50 Milliseconds
Copper Bromide
(578nm)
0.6-1.5mm 200-300 Milliseconds
PDL
(585-600nm)
5.0 - 7.0mm 450-1500 Microseconds
Ruby (Non-Q-Switched)
(694nm)
8 - 10mm Millisecond Range
(Varies w/Mfgr.)
LPIR
(755nm)
7.0 - 12.0mm*** 10 Milliseconds****
Diode
(800/810nm)
1.0 - 10mm Millisecond Range
(Varies w/Mfgr.)
LPIR(Yag.)
(1064nm)
4 - 6mm 50 - 250 millisecond range

* Usual focal distance for handpiece is 5/8 inch
** Usual focal distance for handpiece is 1 1/4 inches
*** Use 10mm Spot for Fluences of 20 J/cm2 or less; use 7mm Spot for fluences > 20Jcm2
**** Begin w/10ms; however, 5 and 20 may be options; double pulsing is suggested for most vessels

# The pulsed mode (0.2 - 0.5 Seconds/Pulse) is appropriate to begin treatment. As the operator becomes facile with the slow, steady movement of the laser handpiece over the outer surfaces of the devices, the use of the continuous beam setting is appropriate. The spot size should be approximately the same diameter as the targeted lesion.

(N.B. When employing an Argon Pumped Tunable Dye Laser, since spot size increases with increased distance between the distal end of the laser handpiece and the target, and since CLO® Devices cause that distance to be double the distance relative to the standard use of this laser, the spot size delivered to the skin surface during treatment will be approximately double the spot size normally delivered by the laser.)


Contact CustomChill